Título : RunScribe Sacral Gait Lab™ Validation for Measuring Pelvic Kinematics during Human Locomotion at Different Speeds
Autor: Ruiz-Malagón, Emilio J. ORCID RESEARCHERID SCOPUSID
García-Pinillos, Felipe ORCID RESEARCHERID SCOPUSID
Molina-Molina, Alejandro ORCID RESEARCHERID SCOPUSID
Soto-Hermoso, Víctor Manuel ORCID RESEARCHERID SCOPUSID
Ruiz-Alias, Santiago ORCID SCOPUSID
Palabras clave : Sacrum; Vertical oscillation; Tilt; Obliquity; Rotation
Fecha de publicación: 27-feb-2023
Editorial : MDPI
Citación : Ruiz-Malagón, E.J.; García-Pinillos, F.; Molina-Molina, A.; Soto-Hermoso, V.M.; Ruiz-Alias, S.A. RunScribe Sacral Gait Lab™ Validation for Measuring Pelvic Kinematics during Human Locomotion at Different Speeds. Sensors 2023, 23, 2604. https:// doi.org/10.3390/s23052604
Resumen : Optoelectronic motion capture systems are considered the gold standard for measuring walking and running kinematics parameters. However, these systems prerequisites are not feasible for practitioners as they entail a laboratory environment and time to process and calculate the data. Therefore, this study aims to evaluate the validity of the three-sensor RunScribe Sacral Gait Lab™ inertial measurement unit (IMU) in measuring pelvic kinematics in terms of vertical oscillation, tilt, obliquity, rotational range of motion, and the maximum angular rates during walking and running on a treadmill. Pelvic kinematic parameters were measured simultaneously using an eight-camera motion analysis system (Qualisys Medical AB, GÖTEBORG, Sweden) and the three-sensor RunScribe Sacral Gait Lab™ (Scribe Lab. Inc. San Francisco, CA, USA) in a sample of 16 healthy young adults. An acceptable level of agreement was considered if the following criteria were met: low bias and SEE (<0.2 times the between-subject differences SD), almost perfect (r > 0.90), and good reliability (ICC > 0.81). The results obtained reveal that the three-sensor RunScribe Sacral Gait Lab™ IMU did not reach the validity criteria established for any of the variables and velocities tested. The results obtained therefore show significant differences between the systems for the pelvic kinematic parameters measured during both walking and running.
URI : https://repositorio.usj.es/handle/123456789/1015
ISSN : 1424-8220
Aparece en las colecciones: Artículos de revistas

Ficheros en este ítem:
Fichero Descripción Tamaño Formato  
RunScribe Sacral Gait Lab™.pdf658,31 kBAdobe PDFVista previa
Visualizar/Abrir


Este ítem está sujeto a una licencia Creative Commons Licencia Creative Commons Creative Commons